Open Circuits and Loss of Deep Brain Stimulation Benefit in the Setting of Parkinson Disease
Open Circuits and Loss of Deep Brain Stimulation Benefit in the Setting of Parkinson Disease
Where did the research take place?
The study site has not been established. Author addresses may differ from where the research occurred.
A plain-language reading has not been prepared for this paper yet.
Original abstract
This case details a patient with PD who underwent bilateral STN DBS for refractory tremor who initially responded well to surgery with a bipolar programming configuration. Seven years after the initial surgery, the patient developed recurrent tremor symptoms and was found to have an open circuit at an active contact. Attempting to use similar settings at an adjacent contact did not sufficiently recapture benefit. While revision surgery was being considered, thorough re-evaluation of functioning contacts demonstrated benefit of tremor control when switching back to a monopolar configuration. The patient has had sustained benefit since that time (~2 years) without requiring a hardware revision. The case demonstrates that in the event of a stable hardware abnormality such as an open circuit, careful reprogramming can be used in select cases in an effort to provide sustained clinical benefit without the need for invasive or costly revision surgery.