RESEARCH / DISCOVERY
← Back to the library

Drop on fixed target reaction initiation approach for serial and time resolved crystallography

Drop on fixed target reaction initiation approach for serial and time resolved crystallography

Read the original publication

Where did the research take place?

The study site has not been established. Author addresses may differ from where the research occurred.

Explore research worldwide

A plain-language reading has not been prepared for this paper yet.

Original abstract

Abstract We describe the design and implementation of a drop on fixed target method for time-resolved serial crystallography at both synchrotron and XFEL facilities. A piezoelectric droplet dispensing pipette is employed for addition of picolitre volume (40 – 90 pL) aqueous droplets, containing (co-)substrate(s) or ligand(s), onto enzyme microcrystals immobilised on a solid support. The system was tested with various enzyme systems, including lysozyme and two ꞵ-lactamases, CTX-M-15 and AmpC EC . Mitigation strategies for cross-well contamination, including the implementation of interleaved controls, are described; the overall performance of the system at synchrotron and X-ray free electron laser facilities was evaluated. This drop on fixed target method is a reliable framework for time-resolved crystallography and will improve the consistency of measurements across facilities.