RESEARCH / DISCOVERY
← Back to the library

The Advantage of Microelectrode Recording When Pneumocephalus Threatens the Precise Placement of a Deep Brain Stimulator.

The Advantage of Microelectrode Recording When Pneumocephalus Threatens the Precise Placement of a Deep Brain Stimulator.

Read the original publication

Where did the research take place?

The study site has not been established. Author addresses may differ from where the research occurred.

Norman, US · Author affiliation

Department of Neurology, Norman Fixel Institute for Neurological Diseases, University of Florida, Gainesville, Florida, United States.
Location evidence

Medellín, CO · Author affiliation

Department of Neurology, Colombian Neurological Institute, Medellin, Colombia.
Location evidence

US · Author affiliation · country only

Department of Neurosurgery, University of Florida, Gainesville, Florida, United States.
Location evidence

Explore research worldwide

A plain-language reading has not been prepared for this paper yet.

Original abstract

CLINICAL VIGNETTE: A 69-year-old woman with Parkinson's disease underwent left subthalamic nucleus (STN) deep brain stimulation (DBS). Intraoperative awake microelectrode recording (MER) was used to confirm targeting. CLINICAL DILEMMA: MER and stimulation mapping revealed a short STN segment on the central pass, absent STN activity on the lateral pass, and low thresholds for capsular side effects. The data suggested a mismatch between the planned imaging-based trajectory and the localization of STN using physiology. CLINICAL SOLUTION: A substantial adjustment based on MER was required, giving up the 'fork' in the brain. The lead was repositioned 3.4 mm posterior and 2.9 mm medial to the initial central pass (4.9 mm vector). Final placement produced robust motor benefit and a desirable therapeutic window for programming. GAP IN KNOWLEDGE: Asleep image-guided workflows assume static intracranial anatomy: pneumocephalus can induce millimetric brain shift. This case demonstrated a pneumocephalus-related displacement and how MER, stimulation thresholds, and postoperative atlas-based validation can be employed to correct it. HIGHLIGHTS: This case illustrates how intraoperative pneumocephalus can compromise targeting in deep brain stimulation.Microelectrode recording provided critical confirmation, guided corrective lead adjustments, and safeguarded therapeutic outcomes, emphasizing the value of physiology-based targeting alongside modern imaging techniques.