Assessment of myocardial sympathetic innervation with [123I]mIBG myocardial scintigraphy: A comparative study between an Anger and a ring-configured cadmium-zinc-telluride gamma camera.

Assessment of myocardial sympathetic innervation with [123I]mIBG myocardial scintigraphy: A comparative study between an Anger and a ring-configured cadmium-zinc-telluride gamma camera.

Read the original publication

Where did the research take place?

The study site has not been established. Author addresses may differ from where the research occurred.

Lund, SE · Author affiliation

Clinical Physiology, Department of Clinical Sciences Lund, Lund University, Skåne University Hospital, Lund, Sweden.
Location evidence

Malmö, SE · Author affiliation

Radiation Physics, Department of Hematology, Oncology and Radiation Physics, Skåne University Hospital, Malmö, Sweden.
Location evidence

Explore research worldwide

A plain-language reading has not been prepared for this paper yet.

Original abstract

BACKGROUND: Assessment of myocardial sympathetic innervation with iodine-123-metaiodobenzylguanidine ([123I]mIBG) scintigraphy is conventionally performed by determining the heart-to-mediastinum ratio (HMR) on planar images. This study aimed to investigate the feasibility of determining HMR from reprojected [123I]mIBG planar images derived from tomographic data by a ring-configured cadmium-zinc-telluride (CZT) gamma camera and its agreement with HMR determined from [123I]mIBG planar images by an Anger gamma camera, in the assessment of myocardial sympathetic innervation. METHODS: A total of 105 study participants (suspected idiopathic rapid-eye-movement sleep behavior disorder n = 55, early-stage Parkinson's Disease n = 36, and healthy controls n = 14) who were referred for [123I]mIBG myocardial scintigraphy were included, in addition to eight phantom experiments simulating a range of HMR. HMR was determined from planar and reprojected planar images acquired with an Anger gamma camera and a ring-configured CZT gamma camera, respectively, for all study participants and phantom experiments. Agreement in HMR between the two imaging methods was analyzed in Bland-Altman plots. Intra- and interobserver variability was evaluated for both imaging methods in a subset of images. RESULTS: The agreement between planar and reprojected planar images was good, with a mean difference in HMR of 0.11 (95% limits of agreement -0.16 to 0.38) for the study participants. Intra- and interobserver variability was low and comparable for both imaging methods. Results from the phantom experiments were consistent with those from the study participants. CONCLUSIONS: Determining HMR from reprojected [123I]mIBG myocardial scintigraphy planar images acquired with a ring-configured CZT gamma camera is feasible and shows good agreement with HMR determined from planar images by a conventional Anger gamma camera.