Detection of deep brain stimulation lead position and orientation in patients using magnetoencephalography.

Detection of deep brain stimulation lead position and orientation in patients using magnetoencephalography.

Read the original publication

Where did the research take place?

The study site has not been established. Author addresses may differ from where the research occurred.

Kiel, DE · Author affiliation

Department of Electrical and Information Engineering, Kiel University, Germany; Department of Neurology, Kiel University, Germany. Electronic address: my@tf.uni-kiel.de.
Location evidence

Boston, US · Author affiliation

Boston Scientific Corporation, Marlborough, United States.
Location evidence

Marlborough, US · Author affiliation

Boston Scientific Corporation, Marlborough, United States.
Location evidence

Düsseldorf, DE · Author affiliation

Institute of Clinical Neuroscience and Medical Psychology, Medical Faculty and University Hospital Düsseldorf, Heinrich Heine University Düsseldorf, Germany.
Location evidence

Explore research worldwide

A plain-language reading has not been prepared for this paper yet.

Original abstract

OBJECTIVE: Deep brain stimulation (DBS) programming in patients with directional DBS requires precise lead position and orientation knowledge. Current computed tomography (CT)-based methods expose patients to radiation. This study investigates a non-radiation-based magnetic detection approach using magnetoencephalography (MEG) in four Parkinson's disease patients. METHODS: MEG recordings were performed under omnidirectional and directional electrode configurations. Three patients were measured with individualized head-casts to minimize head movement. Magnetic detection was applied to determine DBS lead's position and orientation, compared with those derived from postoperative CT imaging. RESULTS: Conventional MEG recordings without head-casts achieved lead position and orientation accuracies of up to 17.3 mm and 24.1°. The use of head-casts improved accuracies to 5.8 ± 1.3 mm and 8.8 ± 2.2° at best. Higher mean errors indicate the presence of systematic biases, primarily caused by the MEG system's limited spatial precision. Reduced error variability demonstrates potential for 1-2 mm localization and 2-4° orientation accuracy. CONCLUSIONS: While magnetic lead position detection is inferior to established approaches, DBS lead orientation could be determined with sufficient accuracy for potential clinical use. Advances in MEG technology may offer improvements in spatial precision and detection accuracy. SIGNIFICANCE: This method may serve as a radiation-free alternative to imaging-based approaches.