Asymptomatic cable twisting in a patient with impending Twiddler syndrome detected during deep brain stimulation surgery for Parkinson's disease: A case report.

Asymptomatic cable twisting in a patient with impending Twiddler syndrome detected during deep brain stimulation surgery for Parkinson's disease: A case report.

Read the original publication

Where did the research take place?

The study site has not been established. Author addresses may differ from where the research occurred.

Surakarta, ID · Author affiliation

Department of Neurosurgery, Dr. Moewardi General Hospital, Surakarta, Indonesia.
Location evidence

Fukuoka, JP · Author affiliation

Department of Neurosurgery, Fukuoka University Hospital, Fukuoka, Japan.
Location evidence

Explore research worldwide

A plain-language reading has not been prepared for this paper yet.

Original abstract

BACKGROUND: Deep brain stimulation (DBS) has consistently demonstrated high efficacy and safety in patients with Parkinson's disease. Twiddler's syndrome is a rare occurrence of hardware failure in patients undergoing neuromodulation. We report here a case of subclinical cable twisting jeopardizing Twiddler's syndrome in a patient with Parkinson's disease who underwent DBS surgery targeting the globus pallidus internus (GPI). CASE DESCRIPTION: A 70-year-old woman with a 7-year history of Parkinson's disease refractory to medication was referred to our department for treatment of involuntary movements of the left hand and leg. She underwent right GPI DBS implantation. Left GPI DBS implantation was subsequently planned to manage resting tremors that developed in the right leg after the first surgery at around one year after the first surgery. During a routine check-up before the second surgery, we incidentally detected Twiddler's syndrome. The patient showed no neurological deficits in the left extremities, the same as before right GPI DBS. We performed left GPI DBS concomitantly with the revision of the implantable pulse generator and extension wire. CONCLUSION: Twiddler's syndrome is a rare complication of DBS. Subclinical risk of cable twisting jeopardizing Twiddler's syndrome is rarely detected without clinical indications of hardware failure. Neurosurgeons should be cognizant of and regularly monitor the implanted device in case serious complications occur.