A Comparative Study on the Lysosomal Cation Channel TMEM175 Using Automated Whole-Cell Patch-Clamp, Lysosomal Patch-Clamp, and Solid Supported Membrane-Based Electrophysiology: Functional Characterization and High-Throughput Screening Assay Development.
A Comparative Study on the Lysosomal Cation Channel TMEM175 Using Automated Whole-Cell Patch-Clamp, Lysosomal Patch-Clamp, and Solid Supported Membrane-Based Electrophysiology: Functional Characterization and High-Throughput Screening Assay Development.
Where did the research take place?
The study site has not been established. Author addresses may differ from where the research occurred.
Munich, DE · Author affiliation
Nanion Technologies, Ganghoferstr. 70a, 80339 Munich, Germany.Location evidence
Regensburg, DE · Author affiliation
RIGeL-Regensburg International Graduate School of Life Sciences, University of Regensburg, 93053 Regensburg, Germany.Location evidence
Glasgow, GB · Author affiliation
SB Drug Discovery, West of Scotland Science Park, Glasgow G20 0XA, UK.Location evidence
Cambridge, US · Author affiliation
Cerevel Therapeutics, 222 Jacobs St, Cambridge, MA 02141, USA.Location evidence
A plain-language reading has not been prepared for this paper yet.
Original abstract
The lysosomal cation channel TMEM175 is a Parkinson's disease-related protein and a promising drug target. Unlike whole-cell automated patch-clamp (APC), lysosomal patch-clamp (LPC) facilitates physiological conditions, but is not yet suitable for high-throughput screening (HTS) applications. Here, we apply solid supported membrane-based electrophysiology (SSME), which enables both direct access to lysosomes and high-throughput electrophysiological recordings. In SSME, ion translocation mediated by TMEM175 is stimulated using a concentration gradient at a resting potential of 0 mV. The concentration-dependent K+ response exhibited an I/c curve with two distinct slopes, indicating the existence of two conducting states. We measured H+ fluxes with a permeability ratio of PH/PK = 48,500, which matches literature findings from patch-clamp studies, validating the SSME approach. Additionally, TMEM175 displayed a high pH dependence. Decreasing cytosolic pH inhibited both K+ and H+ conductivity of TMEM175. Conversely, lysosomal pH and pH gradients did not have major effects on TMEM175. Finally, we developed HTS assays for drug screening and evaluated tool compounds (4-AP, Zn as inhibitors; DCPIB, arachidonic acid, SC-79 as enhancers) using SSME and APC. Additionally, we recorded EC50 data for eight blinded TMEM175 enhancers and compared the results across all three assay technologies, including LPC, discussing their advantages and disadvantages.