RESEARCH / DISCOVERY
← Back to the library

Cold Welding of the Synthes Femoral Neck System Encountered at Conversion to Arthroplasty: A Case Report.

Cold Welding of the Synthes Femoral Neck System Encountered at Conversion to Arthroplasty: A Case Report.

Read the original publication

Where did the research take place?

The study site has not been established. Author addresses may differ from where the research occurred.

Explore research worldwide

A plain-language reading has not been prepared for this paper yet.

Original abstract

CASE: The Synthes Femoral Neck System (FNS) is a relatively new device available to the market in 2019 designed to reduce complication and revision rates after femoral neck fracture stabilization. We present a 77-year-old man with Parkinson disease and avascular necrosis after femoral neck fracture initially managed with the FNS. Subsequent challenges in device removal occurred because of its propensity for bony integration, plate and locking screw welding, and screw head recess destruction. CONCLUSION: Surgeons must recognize the need for additional available equipment (e.g., burr or broken screw removal set) for successful extraction when FNS removal is required.

Explore another example or bring your own paper

Pasted text and PDF extraction stay on this computer. The local guide explains terms and surfaces passages; rewriting requires a configured local model. Scanned PDFs need OCR first.

RECORD & PROVENANCE