RESEARCH / DISCOVERY
← Back to the library

Deep Brain Stimulation Generator Failure due to External Defibrillation in a Patient with Essential Tremor.

Deep Brain Stimulation Generator Failure due to External Defibrillation in a Patient with Essential Tremor.

Read the original publication

Where did the research take place?

The study site has not been established. Author addresses may differ from where the research occurred.

US · Author affiliation · country only

Department of Neurological Surgery, George Washington University Hospital, Washington, District of Columbia, USA, gdavis7@gwu.edu.
Location evidence

Explore research worldwide

A plain-language reading has not been prepared for this paper yet.

Original abstract

There exist only two case reports to date of open cardiac defibrillation with deep brain stimulator system (DBS) implantation. We report a 64-year-old male with DBS system in place for essential tremor who underwent cardiac defibrillation after cardiac arrest. Afterwards, his device impedances were all high and his tremor symptoms returned. Both problems resolved with implantation of a new generator and required no changes to the intracranial leads or extension cables. This is significantly different from the two previous reports. One included a significantly different DBS system relying on transcutaneous RF transmission and reported a lesioning effect after cardioversion. The other utilized a modern DBS system but reported damage to the generator and intracranial leads. We report that only the generator sustained damage, and that there were no intracranial changes that occurred.

Explore another example or bring your own paper

Pasted text and PDF extraction stay on this computer. The local guide explains terms and surfaces passages; rewriting requires a configured local model. Scanned PDFs need OCR first.

RECORD & PROVENANCE