RESEARCH / DISCOVERY
← Back to the library

Modeling of Electric Fields in Individual Imaging Atlas for Capsular Threshold Prediction of Deep Brain Stimulation in Parkinson's Disease: A Pilot Study.

Modeling of Electric Fields in Individual Imaging Atlas for Capsular Threshold Prediction of Deep Brain Stimulation in Parkinson's Disease: A Pilot Study.

Read the original publication

Where did the research take place?

The study site has not been established. Author addresses may differ from where the research occurred.

Besançon, FR · Author affiliation

Department of Neurology, Besançon University Hospital, Besançon, France.
Location evidence

Geneva, CH · Author affiliation

Department of Neurology, Geneva University Hospital, Geneva, Switzerland.
Location evidence

Boston, US · Author affiliation

Boston Scientific Corporation, Valencia, CA, United States.
Location evidence

Valencia, US · Author affiliation

Boston Scientific Corporation, Valencia, CA, United States.
Location evidence

Bern, CH · Author affiliation

Department of Neurology, Bern University Hospital, Bern, Switzerland.
Location evidence

Explore research worldwide

A plain-language reading has not been prepared for this paper yet.

Original abstract

Background: Modeling of deep brain stimulation electric fields and anatomy-based software might improve post-operative management of patients with Parkinson's disease (PD) who have benefitted from subthalamic nucleus deep brain stimulation (STN-DBS). Objective: We compared clinical and software-guided determination of the thresholds for current diffusion to the pyramidal tract, the most frequent limiting side effect in post-operative management of STN-DBS PD patients. Methods: We assessed monopolar reviews in 16 consecutive STN-DBS PD patients and retrospectively compared clinical capsular thresholds, which had been assessed according to standard clinical practice, to those predicted by volume of tissue activated (VTA) model software. All the modeling steps were performed blinded from patients' clinical evaluations. Results: At the group level, we found a significant correlation (p = 0.0001) when performing statistical analysis on the z-scored capsular thresholds, but with a low regression coefficient (r = 0.2445). When considering intra-patient analysis, we found significant correlations (p < 0.05) between capsular threshold as modeled with the software and capsular threshold as determined clinically in five patients (31.2%). Conclusions: In this pilot study, the VTA model software was of limited assistance in identifying capsular thresholds for the whole cohort due to a large inter-patient variability. Clinical testing remains the gold standard in selecting stimulation parameters for STN-DBS in PD.