The Unsuccessful Twiddler: A Case of Twiddler's Syndrome Without Deep Brain Stimulator Lead Breakage.
The Unsuccessful Twiddler: A Case of Twiddler's Syndrome Without Deep Brain Stimulator Lead Breakage.
Where did the research take place?
The study site has not been established. Author addresses may differ from where the research occurred.
Moreno Valley, US · Author affiliation
Neurosurgery, Riverside University Health System Medical Center, Moreno Valley, USA.Location evidence
US · Author affiliation · country only
Neurosurgery, University of California Riverside, Riverside, USA.Location evidence
Los Angeles, US · Author affiliation
Neurosurgery, Kaiser Permanente, Los Angeles, USA.Location evidence
A plain-language reading has not been prepared for this paper yet.
Original abstract
The authors present the case of a 78-year-old right-handed female with a past medical history of Parkinson's disease, treated with implantation of a left-sided subthalamic nucleus St. Jude Medical Infinity® (Abbott Medical, Austin, TX) deep brain stimulator, who presented with lead-associated discomfort, or "bowstringing". Further investigation by chest X-ray revealed an extensive case of distal lead coiling. However, it was surprising that, despite the extensive coiling, the lead stayed intact without hardware failure as proven by patient remaining asymptomatic from her Parkinson's disease and intraoperative impedance testing demonstrating normal results. After revision surgery, the patient remained asymptomatic. Due to paucity of cases of this disease in the literature, specific predictive risk factors are not known, but certain patient characteristics may help take precautions.