Inappropriate Shock from Subcutaneous Implantable Cardioverter-Defibrillator Caused by Electrical Remodeling After Wolff-Parkinson-White Ablation.
Inappropriate Shock from Subcutaneous Implantable Cardioverter-Defibrillator Caused by Electrical Remodeling After Wolff-Parkinson-White Ablation.
Where did the research take place?
The study site has not been established. Author addresses may differ from where the research occurred.
Aventura, US · Author affiliation
Internal Medicine, Aventura Hospital and Medical Center, Aventura, USA.Location evidence
Sacramento, US · Author affiliation
Cardiology, UC Davis Medical Center, Sacramento, USA.Location evidence
Kendall, US · Author affiliation
Anesthesiology, Kendall Regional Medical Center, Miami, USA.Location evidence
Miami, US · Author affiliation
Anesthesiology, Kendall Regional Medical Center, Miami, USA.Location evidence
A plain-language reading has not been prepared for this paper yet.
Original abstract
Hypertrophic cardiomyopathy (HCM) and Wolff-Parkinson-White syndrome have been associated with sudden cardiac death. A subcutaneous implantable cardioverter-defibrillator (S-ICD) is an effective device used to reduce the risk of sudden cardiac death in these patients. The most common cause of inappropriate shocks with S-ICD is T-wave oversensing. We present the case of a 19-year-old man with repeated shocks from his S-ICD. This case highlights some of the sensing issues related to the S-ICD that can result in inappropriate shocks. A vector change may have occurred after T-wave remodeling, post accessory pathway ablation, and loss of R-waves due to HCM scar progression, leading to this consequence.