Validity of the wall goniometer as a screening tool to detect postural abnormalities in Parkinson's disease.
Validity of the wall goniometer as a screening tool to detect postural abnormalities in Parkinson's disease.
Where did the research take place?
The study site has not been established. Author addresses may differ from where the research occurred.
Verona, IT · Author affiliation
Neurology Unit, Movement Disorders Division, Department of Neurosciences, Biomedicine and Movement Sciences, University of Verona, Verona, Italy.Location evidence
Turin, IT · Author affiliation
Department of Neuroscience "Rita Levi Montalcini", University of Torino, Torino, Italy.Location evidence
Pisa, IT · Author affiliation
Department of Clinical and Experimental Medicine, University of Pisa, Italy.Location evidence
Ancona, IT · Author affiliation
Department of Experimental and Clinical Medicine, Neurorehabilitation Clinic, "Politecnica delle Marche" University, Ancona, AN, Italy.Location evidence
IT · Author affiliation · country only
Department of Neuroscience, Imaging and Clinical Sciences, University G.d'Annunzio of Chieti-Pescara, Italy.Location evidence
Trieste, IT · Author affiliation
Clinical Neurology Unit, Department of Medical, Surgical and Health services, University of Trieste, Italy.Location evidence
Rome, IT · Author affiliation
University and Institute for Research and Medical Care IRCCS San Raffaele, Roma, Italy.Location evidence
Toronto, CA · Author affiliation
Edmond J. Safra Program in Parkinson's Disease and the Morton and Gloria Shulman Movement Disorders Clinic, Toronto Western Hospital, UHN, Division of Neurology, University of Toronto, Toronto, Ontario, Canada; Krembil Brain Institute, Toronto, Ontario, Canada.Location evidence
A plain-language reading has not been prepared for this paper yet.
Original abstract
INTRODUCTION: Software-based measurements of postural abnormalities in Parkinson's disease (PD) are the gold standard but may be time-consuming and not always feasible in clinical practice. Wall goniometer (WG) is an easier, quicker, and inexpensive instrument for screening patients with postural abnormalities, but no studies have investigated its validity so far. The aim of this study was to investigate the validity of the WG to measure postural abnormalities. METHODS: A total of 283 consecutive PD outpatients with ≥5° forward trunk, lateral trunk or forward neck bending (FTB, LTB, FNB, respectively) were recruited from seven centers for movement disorders. Postural abnormalities were measured in lateral and posterior view using a freeware program (gold standard) and the WG. Both angles were expressed in degrees (°). Sensitivity and specificity for the diagnosis of camptocormia, Pisa syndrome, and anterocollis were assessed. RESULTS: WG showed good to excellent agreement (intraclass correlation coefficient from 0.80 to 0.98) compared to the gold standard. Bland-Altman plots showed a mean difference between the methods from -7.4° to 0.4° with limits of agreements from -17.7° to 9.5°. Sensitivity was 100% for the diagnosis of Pisa syndrome, 95.74% for anterocollis, 76.67% for upper camptocormia, and 63.64% for lower camptocormia. Specificity was 59.57% for Pisa syndrome, 71.43% for anterocollis, 89.80% for upper camptocormia, and 100% for lower camptocormia. Overall, the WG underestimated measurements, especially in lower camptocormia with an average of -8.7° (90% of cases). CONCLUSION: WG is a valid tool for screening Pisa syndrome and anterocollis, but approximately 10° more should be added for camptocormia.