Original abstract
The need for cardiac device implantation in patients receiving deep brain stimulation (DBS) is increasing. Despite the theoretical risk of the two systems interacting, there are no clear guidelines for cardiologists carrying out cardiac device implantation in this population. We performed a review of the literature and describe 13 case reports in which patients have both DBS and a cardiac pacemaker or ICD implanted. Except for one early study, in which an ICD shock reset the deep brain stimulator, no significant interactions have been reported. We discuss the potential interactions between DBS and cardiac devices, and provide practical advice for implanting cardiologists. We conclude that, provided that specific precautions are taken, cardiac device implantation is likely to be safe in patients with DBS.