RESEARCH / DISCOVERY
← Back to the library

Explant Analysis from a Patient Exhibiting Rapid Acceleration of Parkinson Disease Symptoms and Hypercobaltemia Following Metal-on-Metal Total Hip Arthroplasty: A Case Report.

Explant Analysis from a Patient Exhibiting Rapid Acceleration of Parkinson Disease Symptoms and Hypercobaltemia Following Metal-on-Metal Total Hip Arthroplasty: A Case Report.

Read the original publication

Where did the research take place?

The study site has not been established. Author addresses may differ from where the research occurred.

Explore research worldwide

A plain-language reading has not been prepared for this paper yet.

Original abstract

CASE: A 46-year-old man developed rapidly progressing Parkinson disease symptoms after metal-on-metal total hip arthroplasty, worsening to the extent that he underwent implantation of a right subthalamic deep brain stimulator (DBS). A serum cobalt level of 116 μg/L was measured shortly after DBS implantation, and revision arthroplasty was performed using ceramic-on-polyethylene bearings. After revision, the patient's serum cobalt level fell below 1 μg/L, the Parkinson disease symptoms improved, and he reduced the DBS to its lowest setting. Explant analysis showed substantial material loss from the bearing surfaces of one hip. CONCLUSION: Future studies should evaluate the possible link between Parkinson disease symptoms and prosthesis wear-induced hypercobaltemia.