Clinical Tremor Severity Estimation Using an Instrumented Eating Utensil.
Clinical Tremor Severity Estimation Using an Instrumented Eating Utensil.
Where did the research take place?
The study site has not been established. Author addresses may differ from where the research occurred.
San Francisco, US · Author affiliation
Department of Neurology, University of California San Francisco, CA, USA.Location evidence
Mountain View, US · Author affiliation
Verily Life Sciences, Mountain View, CA, USA.Location evidence
A plain-language reading has not been prepared for this paper yet.
Original abstract
We demonstrate the feasibility of estimating clinical tremor scores using an eating utensil with motion-sensing and tremor-cancellation technology in thirteen patients with tremor. Three experts scored hand tremor using the modified Fahn- Tolosa-Marin (FTM) scale. A linear model was trained to estimate tremor severity using the recorded motion signals. The average neurologist FTM score was 1.6±0.7 for PD and 2.6±0.7 for ET patients. The average model score was 1.6±0.7 for PD and 2.6±0.6 for ET. Correlation coefficient between the clinical and model tremor scores was 0.91 (p < 0.001). Motion data from an instrumented eating utensil accurately derived tremor ratings enabling practical, objective daily monitoring.