Comparative study of microelectrode recording-based STN location and MRI-based STN location in low to ultra-high field (7.0 T) T2-weighted MRI images.

Comparative study of microelectrode recording-based STN location and MRI-based STN location in low to ultra-high field (7.0 T) T2-weighted MRI images.

Read the original publication

Where did the research take place?

The study site has not been established. Author addresses may differ from where the research occurred.

Amsterdam, NL · Author affiliation

Department of Neurology and Clinical Neurophysiology, Academic Medical Center, Amsterdam, The Netherlands.
Location evidence

NL · Author affiliation · country only

Department of Neurology, Haga Teaching Hospital, Den Haag, The Netherlands.
Location evidence

Leiden, NL · Author affiliation

Department of Neurology, Leiden University Medical Center, Leiden, The Netherlands.
Location evidence

Explore research worldwide

A plain-language reading has not been prepared for this paper yet.

Original abstract

OBJECTIVE: The correspondence between the anatomical STN and the STN observed in T2-weighted MRI images used for deep brain stimulation (DBS) targeting remains unclear. Using a new method, we compared the STN borders seen on MRI images with those estimated by intraoperative microelectrode recordings (MER). APPROACH: We developed a method to automatically generate a detailed estimation of STN shape and the location of its borders, based on multiple-channel MER measurements. In 33 STNs of 19 Parkinson patients, we quantitatively compared the dorsal and lateral borders of this MER-based STN model with the STN borders visualized by 1.5 T (n = 14), 3.0 T (n = 10) and 7.0 T (n = 9) T2-weighted MRI. MAIN RESULTS: The dorsal border was identified more dorsally on coronal T2 MRI than by the MER-based STN model, with a significant difference in the 3.0 T (range 0.97-1.19 mm) and 7.0 T (range 1.23-1.25 mm) groups. The lateral border was significantly more medial on 1.5 T (mean: 1.97 mm) and 3.0 T (mean: 2.49 mm) MRI than in the MER-based STN; a difference that was not found in the 7.0 T group. SIGNIFICANCE: The STN extends further in the dorsal direction on coronal T2 MRI images than is measured by MER. Increasing MRI field strength to 3.0 T or 7.0 T yields similar discrepancies between MER and MRI at the dorsal STN border. In contrast, increasing MRI field strength to 7.0 T may be useful for identification of the lateral STN border and thereby improve DBS targeting.