When battery exhaustion lets the lame walk: a case report on the importance of long-term stimulator monitoring in deep brain stimulation.
When battery exhaustion lets the lame walk: a case report on the importance of long-term stimulator monitoring in deep brain stimulation.
Where did the research take place?
The study site has not been established. Author addresses may differ from where the research occurred.
Göttingen, DE · Author affiliation
Department of Clinical Neurophysiology, University Medical Centre, Göttingen, Robert-Koch-Str. 40, D-37075, Göttingen, Germany. msommer@gwdg.de.Location evidence
A plain-language reading has not been prepared for this paper yet.
Original abstract
BackgroundDeep brain stimulation is increasingly used in the treatment of advanced Parkinson's disease. While its short-term effectiveness is well documented, there are only few reports on long-term outcomes, and the need to repeatedly reprogram the stimulator is seldom reported.Case presentationWe present a 74-year-old man with gait impairment, which had been mistaken for worsening of the disease and only remitted when the stimulator battery was exhausted indicating that the stimulator itself had been the cause.ConclusionThis case highlights the need to repeatedly monitor not only battery capacity, but also stimulator-related side-effects for an extended period after implantation and, if necessary, to refer to centres capable of systematically reprogramming the device.