Fabrication and initial testing of the μDBS: a novel Deep Brain Stimulation electrode with thousands of individually controllable contacts.
Fabrication and initial testing of the μDBS: a novel Deep Brain Stimulation electrode with thousands of individually controllable contacts.
Where did the research take place?
The study site has not been established. Author addresses may differ from where the research occurred.
Salt Lake City, US · Author affiliation
University of Utah, Salt Lake City, UT, 84112, USA, andrew.willsie@utah.edu.Location evidence
A plain-language reading has not been prepared for this paper yet.
Original abstract
High frequency electrical stimulation of deep brain structures such as the subthalamic nucleus in Parkinson's disease or thalamus for essential tremor is used clinically to reduce symptom severity. Deep brain stimulation activates neurons in specific brain structures and connection pathways, overriding aberrant neural activity associated with symptoms. While optimal deep brain stimulation might activate a particular neural structure precisely, existing deep brain stimulation can only generate roughly-spherical regions of activation that do not overlap with any target anatomy. Additionally, side effects linked to stimulation may be the result of limited control over placement of stimulation and its subsequent spread out of optimal target boundaries. We propose a novel lead with thousands of individually controllable contacts capable of asymmetric stimulation profiles. Here we outline the design motivation, manufacturing process, and initial testing of this new electrode design, placing it on track for further directional stimulation studies.