Signal features of surface electromyography in advanced Parkinson's disease during different settings of deep brain stimulation.
Signal features of surface electromyography in advanced Parkinson's disease during different settings of deep brain stimulation.
Where did the research take place?
The study site has not been established. Author addresses may differ from where the research occurred.
FI · Author affiliation · country only
Department of Applied Physics, University of Eastern Finland, Finland. Electronic address: saara.rissanen@uef.fi.Location evidence
Helsinki, FI · Author affiliation
Department of Neurology, Helsinki University Central Hospital, Finland; BioMag Laboratory, Helsinki University Central Hospital, Finland.Location evidence
Tampere, FI · Author affiliation
Department of Physical and Rehabilitation Medicine, Tampere University Hospital, Finland.Location evidence
Kuopio, FI · Author affiliation
Department of Physical and Rehabilitation Medicine, Kuopio University Hospital, Finland.Location evidence
A plain-language reading has not been prepared for this paper yet.
Original abstract
ObjectiveElectromyography (EMG) and acceleration (ACC) measurements are potential methods for quantifying efficacy of deep brain stimulation (DBS) treatment in Parkinson's disease (PD). The treatment efficacy depends on the settings of DBS parameters (pulse amplitude, frequency and width). This study quantified, if EMG and ACC signal features differ between different DBS settings and if DBS effect is unequal between different muscles.MethodsEMGs were measured from biceps brachii (BB) and tibialis anterior (TA) muscles of 13 PD patients. ACCs were measured from wrists. Measurements were performed during seven different settings of DBS and analyzed using methods based on spectral analysis, signal morphology and nonlinear dynamics.ResultsThe results showed significant within-subject differences in the EMG signal kurtosis, correlation dimension, recurrence rate and EMG-ACC coherence between different DBS settings for BB but not for TA muscles. Correlations between EMG feature values and clinical rest tremor and rigidity scores were weak but significant.ConclusionsSurface EMG features differed between different DBS settings and DBS effect was unequal between upper and lower limb muscles.SignificanceEMG changes pointed to previously defined optimal settings in most of patients, which should be quantified even more deeply in the upcoming studies.